Hospira Issues a Voluntary Nationwide Recall for Two Lots of Naloxone Hydrochloride Injection, USP, in the Carpuject™ Syringe System due to the Potential Presence of Particulate Matter

Hospira, Inc., a Pfizer company, is voluntarily recalling lots 72680LL and 76510LL of Naloxone Hydrochloride Injection, USP, 0.4 mg/mL, 1 mL in 2.5 mL, Carpuject Single-use cartridge syringe system (NDC 0409-1782-69), to the hospital/institution level due to the potential presence of embedded and loose particulate matter on the syringe plunger.

SunMed Holdings, LLC Issues a Nationwide Recall of STAT-Check and Medline Manual Resuscitator Bags

On May 29, 2018, SunMed Holdings, LLC initiated a nationwide recall of 18,808 units of STAT-Check and Medline resuscitator bags which were distributed between February 1, 2018 to May 13, 2018. The patient port retaining ring of the affected bags may not fully seat which may allow the patient port to detach during use. If this takes place, the resuscitator bags may not deliver air to the patient and result in a delay in treatment and life-threatening health consequences. There have been no reported injuries to date.

Abbott Implements Corrective Action for Heartmate 3 Heart Pump

Abbott is communicating to physicians about a field corrective action related to the HeartMate 3™ Left Ventricular Assist Device (LVAD). Abbott notified physicians and global regulatory bodies on April 5, 2018, that outflow graft twisting could occur post-implant in the HeartMate 3 LVAD, resulting in a persistent low flow alarm that may signal a potential safety risk to patients, such as low blood flow or clotting.